Calibration – American Magnetics 185 & 186 Liquid Level Instruments (CE-Marked) User Manual

Page 29

Advertising
background image

Rev. 3

19

Calibration

Calibration methods

3.3 Calibration Methods for Model 185/186 Instruments

The most straightforward calibration method is the Open Dewar
Calibration

which requires the customer to have access to a filled dewar

where the full active length of the sensor can be dipped. The Closed Dewar
Calibration

method can be performed in situations where it is not feasible

for the customer to dip the sensor into an open dewar, such as situations

where the target liquid is under pressure. The closed dewar calibration is

more complex and may require initial preparations to insure success.

Occasionally customers ask AMI to calibrate an instrument and sensor for

a liquid which is not available at AMI for calibration purposes and/or for a

sensor which is too long to be calibrated at our facilities.

For the case of the target liquid being unavailable, AMI uses liquid

nitrogen as the reference liquid and an Approximate Calibration is

performed using mathematical manipulation of the ratio of the dielectric

constants between liquid nitrogen and the desired liquid. This procedure is

outlined in the Approximate Calibration section beginning on page 25. The

technique is intended to provide the instrument with an approximate

calibration so that it can be used immediately by the customer. However,

the customer is still expected to perform a more accurate calibration where

feasible, such as the open dewar or closed dewar calibration, with the

target liquid.

For the case where a sensor is too long to be calibrated in AMI facilities,

AMI will perform a partial length open dewar calibration in liquid

nitrogen, and then calculate the MAX calibration point. A dielectric ratio

may also be subsequently utilized to adjust for a target liquid other than

liquid nitrogen. The customer is expected to perform a more accurate open

dewar or closed dewar calibration if feasible.

As a quick guide for selection of the best calibration method available, a

calibration selection diagram is presented below. If the instrument and

sensor are purchased as a unit from AMI, then the factory calibration will

be adequate in most cases. However, for the exceptions noted in the

previous paragraphs (which are approximate calibrations), the customer

should perform a more accurate open dewar or closed dewar calibration. A

customer performed calibration is also required for sensors that are

purchased as a separate item from the instrument, since the instrument

and sensor were not both available for calibration at AMI facilities.

Advertising