3 approximate calibration, Calibration – American Magnetics 185 & 186 Liquid Level Instruments (CE-Marked) User Manual

Page 35

Advertising
background image

Rev. 3

25

Calibration

Approximate calibration

calibrated with a standard active region of 20". The LENGTH

setting of the instrument should also be configured for 20".

7. Proceed to the Operation section for directions for configuring the

instrument.

3.3.3 Approximate calibration

This procedure is the least accurate form of calibration and should be used

only when the aforementioned calibration procedures are not viable. The

approximate calibration method can be used in cases where the sensor

cannot be dipped into the target liquid, the full active length of the sensor

cannot be dipped into an open dewar, or both. Approximate calibration

may also be useful for situations where the sensor cannot be dipped into

the target liquid under the expected operating pressure.

If the target liquid is not available for dipping, a substitute non-conducting

reference liquid can be used. If the full length of the sensor cannot be

dipped, then a partial length dip can be performed. If both situations are

encountered, then a partial length dip can be performed in a substitute

reference liquid.

1. First, cool the sensor as much as possible by dipping the sensor as

far as possible in the available reference liquid.

2. Slowly withdraw the sensor out of the reference liquid until the

level is approximately even with the bottom hole in the sensor and

then press the MIN push-button through the small hole provided in

the instrument front panel. When the calibration point has been

accepted, the display will show "bbb.b" and the push-button can

then be released. The location of the liquid level on the sensor when

the MIN button is pressed becomes the 0% level.

Note

Having a small amount of liquid in contact with the sensor at the
MIN calibration level helps stabilize the sensor capacitance for 0%
level indication.

3. Reinsert the sensor in the reference liquid as far as possible, but

not exceeding 1" below the top hole. Note the physical location of

the liquid level on the sensor at the maximum insertion depth.

While the sensor is submerged at the maximum depth, press the

MAX push-button through the small hole provided in the

instrument front panel. When the calibration point has been

accepted, the display will show "bbb.b" and the push-button can

then be released.

Advertising