2 sensor contamination, Operation – American Magnetics 185 & 186 Liquid Level Instruments (CE-Marked) User Manual

Page 44

Advertising
background image

34

Rev. 3

Operation

Sensor contamination

4.2 Sensor Contamination

To ensure proper instrument calibration and operation, care must be

taken to ensure the sensor is kept free of contaminants and not subjected

to any force which would physically distort the sensor. Water or other

electrically conducting substances in the sensor will disturb the measured

capacitance and therefore instrument response. Physically distorting the

sensor in any way will also cause abnormal instrument operation by

introducing variations in the sensor capacitance not due to liquid level.

The absolute calibration of the instrument can be inaccurate if care is not

taken to ensure the sensor is in a proper environment.

Cold sensors exposed to humidified air can show erroneous high level

readings due to the fact that the air contains moisture which can condense

between the cold sensing tubes. A small film of water can cause a shorted

or partially shorted condition, which results in false level readings. As the

sensor warms, the moisture may evaporate and the sensor will again read

correctly. This is a physical phenomenon and does not indicate any

problem with your AMI level equipment. Limit or eliminate exposure of

cold sensors to humidified air to avoid this condition.

If a sensor should require cleaning, flushing with alcohol is recommended.

The sensor cannot be used again until all the alcohol has been evaporated.

Under no circumstances should the sensor be disassembled.

Advertising