6 commands for performing remote calibration, 5 serial poll status byte, Remote interface reference – American Magnetics 185 & 186 Liquid Level Instruments (CE-Marked) User Manual

Page 59

Advertising
background image

Rev. 3

49

Remote Interface Reference

Serial Poll Status Byte

5.2.4.6 Commands for performing remote calibration

The calibration commands perform a remote calibration equivalent to

activating the front panel MIN, MAX, and “~” (approximate) calibration

buttons. The calibration is automatically saved to permanent memory. See

the Calibration section for more information regarding calibration.

5.2.5

Serial poll status byte

The IEEE-488 serial poll status byte (spoll byte) can be used to obtain

information about the state of the instrument. Bit 7 of the status byte is

reserved for SRQ. The remaining bits are used to provide custom

information as shown in the table below.

Note

The fill mode indication is only accurate if the fill mode toggle
switch on the front panel is in the AUTO position. There is no
remote indication or control available for the OPEN or CLOSED
manual override selections.

Command: MINCAL

Function: Performs a MIN

calibration

Returns: MINCAL

Command: MAXCAL

Function: Performs a MAX

calibration

Returns: MAXCAL

Command: APPROX=

<value>

Function: Performs an

approximate

calibration using

<

value>

as the

approximate

calibration factor

Returns: APPROX =

<value>

Bit

ON

OFF

1 (LSB)

HI relay on

HI relay off

2

A relay on

A relay off

3

B relay on

B relay off

4

LO relay on

LO relay off

5

Fill mode on

(controller output energized)

Fill mode off

(controller output de-

energized)

6

Data ready

No data available

7

Service Request (SRQ)

No SRQ

8

Not used

Not used