Calibration – American Magnetics 185 & 186 Liquid Level Instruments (CE-Marked) User Manual

Page 36

Advertising
background image

26

Rev. 3

Calibration

Approximate calibration

4. Measure the distance between the bottom hole of the sensor and

the location of the liquid level noted during step 3. This measured

length is L

dipped

.

5. The dielectric constant for the reference liquid, e

1

, and the target

liquid, e

2

, must be known to complete the approximate calibration.

These values must be placed in the equation:

where L

dipped

is the length of the sensor dipped in the reference

liquid and L

active

is the active sensor length.

Note

If the target liquid is available for dipping (i.e. the reference liquid
and target liquid are the same), then the dielectric ratio,

(e

2

1)/(e

1

1), becomes 1. If the full active length of the sensor can be dipped,
then the length ratio, L

active

/ L

dipped

, becomes

1.

Note that e

1

= 1.454 for liquid nitrogen at

203

°

C at atmospheric

pressure. Dielectric constants for several liquids are provided in the
Appendix

. The dielectric constant varies with temperature and

pressure, therefore for best accuracy use the dielectric constant for

the target liquid at the temperature and pressure maintained in

the containing vessel.

6. Once the approximate calibration factor is calculated, it can be

entered into the instrument by placing the front panel control mode

rotary switch in the SILENCE position. By using the RAISE/

LOWER toggle switch and holding it in the up or down position,

you can adjust the approximate calibration factor up or down. The

display will move slowly at first and then faster. Once near the

desired value, simply release the switch momentarily and then

resume changing the factor at the slower speed. Once the desired

number has been reached, release the toggle switch.

7. Once the approximate calibration factor has been entered,

momentarily press the CAL push-button labeled as "~" (the tilde

character) through the small hole provided in the instrument front

panel. When the calibration factor has been accepted, the display

will show "ddd.d" and the button can then be released. This

completes the approximate calibration procedure.

Approximate Calibration Factor

e

2

1

e

1

1

--------------

100

×

L

active

L

dipped

-----------------

=

Advertising