Figure 3-2. calibration method selection diagram, Calibration, Rev. 3 – American Magnetics 185 & 186 Liquid Level Instruments (CE-Marked) User Manual

Page 30

Advertising
background image

20

Rev. 3

Calibration

Calibration methods

Figure 3-2.

Calibration method selection diagram.

Instrument &

sensor purchased

together?

START

Verify sensor

specifications

Verify calibration

sticker on bottom of

instrument

Perform

Open

Dewar Calibration

Is factory

calibration

correct?

Can

Open

Dewar Calibration

be

performed in target

liquid?

Can

Closed

Dewar Calibration

be

performed in target

liquid?

Perform

Closed

Dewar Calibration

Perform

Approximate

Calibration

Go to Operation

Y

N

N

Y

N

Is factory

calibration

Approximate?

Y

Y**

N

** Factory Approximate Calibration should only be used if

open dewar or closed dewar calibration is not feasible in

the target liquid at the customer’s facility.

N

Can

Open

Dewar Calibration

be

performed in target

liquid?

Can

Closed

Dewar Calibration

be

performed in target

liquid?

Y

Y

N

N

Y