Figure 3-2. calibration method selection diagram, Calibration, Rev. 3 – American Magnetics 185 & 186 Liquid Level Instruments (CE-Marked) User Manual
Page 30
20
Rev. 3
Calibration
Calibration methods
Figure 3-2.
Calibration method selection diagram.
Instrument &
sensor purchased
together?
START
Verify sensor
specifications
Verify calibration
sticker on bottom of
instrument
Perform
Open
Dewar Calibration
Is factory
calibration
correct?
Can
Open
Dewar Calibration
be
performed in target
liquid?
Can
Closed
Dewar Calibration
be
performed in target
liquid?
Perform
Closed
Dewar Calibration
Perform
Approximate
Calibration
Go to Operation
Y
N
N
Y
N
Is factory
calibration
Approximate?
Y
Y**
N
** Factory Approximate Calibration should only be used if
open dewar or closed dewar calibration is not feasible in
the target liquid at the customer’s facility.
N
Can
Open
Dewar Calibration
be
performed in target
liquid?
Can
Closed
Dewar Calibration
be
performed in target
liquid?
Y
Y
N
N
Y