A.1 4-20 mA current loop option

The 4-20 mA output utilizes pins 1 and 2 of connector J2. When the Model

185/186 is configured for the 4-20 mA current loop option, the 0-10 VDC

analog output from connector J2 is not available. The figure below shows

the wiring diagram and the voltage requirements for the power supply and

receiver.

Caution

It is extremely important to observe all polarities and to not exceed
+32 VDC for the loop power supply in order to prevent damage to the
4-20 mA driver circuit.

Note

For maximum immunity to external electrical and electromagnetic
disturbances, all external cabling (except for the AC input, controller
output, and coaxial cabling) should be shielded. The cable shield
should be connected to the chassis of the instrument by connecting to
the D-sub connector shell.

Receiver

AMI

Level

Instrument

Loop

Power

Supply

+13 to +32 VDC

R

L

I

LOOP

J2 pin 1 (+)

+

+

_

_

J2 pin 2 (-)

J2 pin 4 (common)