7 select the appropriate units display option, 8 connect the optional analog output signal, Operation – American Magnetics 185 & 186 Liquid Level Instruments (CE-Marked) User Manual

Page 43

Advertising
background image

Rev. 3

33

Operation

Fill timer (INTERVAL)

Note

The INTERVAL function is disabled when the INTERVAL setting is
“0.0”. Adjusting the INTERVAL setting to “0.0” will also terminate
any in-progress functions of the INTERVAL timer.

The INTERVAL setting can be adjusted by placing the control mode rotary

switch in the INTERVAL position and using the RAISE/LOWER toggle

switch to adjust the setting up or down. The display will move slowly at

first and then faster. Once near the desired value which is displayed in

minutes, simply release the switch momentarily and then resume

changing the setpoint at the slower speed. The instrument is shipped from

the factory with a zero interval time.

4.1.7

Select the appropriate units display option

Place the units mode toggle switch in the position desired for the display

output units during operation. The % position displays the percentage of

active sensor length that is immersed in liquid.

4.1.8

Connect the optional analog output signal

If the instrument was purchased with an analog output option, the

auxiliary connector J2 on the rear of the instrument provides a 0-10 VDC

or 4-20 mA analog signal corresponding to 0-100% of liquid level. Refer to

the Appendix for the pinout diagram of connector J2.