IEEE-488 Communication Option

5.2 IEEE-488 Communication Option

The IEEE-488 communication option provides a GPIB connector on the

rear panel of the instrument for IEEE-488 (GPIB, HPIB) communications.

5.2.1

Command/return termination characters

All commands are transmitted and received as ASCII values and are case

insensitive. The Model 185/186 always transmits <LF> and EOI as the

termination for return data. The Model 185/186 can accept <CR>, <LF>,
<CR><LF>, <LF><CR>,

or <LF> with EIO as termination characters from

an external IEEE-488 interface.

Only one command at a time should be sent to the Model 185/186 by the

external IEEE-488 interface. Additional commands separated by a

semicolon will not be processed. The instrument uses a single 16 character

buffer for input and output. Consequently, all input strings including

terminations should not be longer than 16 characters. Any excess

characters will be discarded. All alphabetical characters are case

insensitive and character encoding is in accordance with IEEE 488.2.

5.2.2

Communicating with the Model 185/186

The use of a single buffer for both input and output is a result of memory

limitations in the Model 185/186. In order to keep the external IEEE-488

interface from sending successive commands faster than the Model 185/

186 can respond, the Model 185/186 uses the Serial Poll Service Request

(SRQ) to let the external computer know it has finished processing the last

command received and is ready to send a response. This is true of all

commands. Thus sending commands to the Model 185/186 using IEEE-488

protocol is a three step process: 1) send the ASCII command, 2) wait for

SRQ, and 3) get the instrument response.

Note

API’s for some manufacturer’s cards require you to use different
functions to check for SRQ and read the serial poll status (spoll)
byte. Invoking the command to read the spoll byte may be required
to actually clear the SRQ status.

A basic flow diagram for sending an ASCII command to the Model 185/186

and receiving a response is shown on the following page. It is not necessary

to wait exclusively for the SRQ status from the instrument. Other bus

commands can be processed while waiting for the SRQ status from the

instrument.