2 connecting the profibus cable, Connecting the profibus cable -2 – YSI DIQ/S 182-PR System 182 User Manual

Page 112

Advertising
background image

PROFIBUS connection

System 182

7 - 2

ba76024e01

02/2012

 Have admissible bus parameters been set only? If you are in doubt

set them to default values.

 Is it guaranteed that the PROFIBUS master has consistent access

to the 16 byte sensor data structure?

Note for Siemens PLC with programming language STEP7:
The following commands have to be used for consistent access:

– SFC15 (for writing)

– SFC14 (for reading)

 After commissioning:

Do all devices signal error-free behavior?
(DIQ/S 182-PR: status "Online" in the measured value and status
display).

7.2

Connecting the PROFIBUS cable

The PROFIBUS cable is connected to the DIQ/S 182-PR using a 9-pin
D-SUB connector on the top of the housing. As connector on the cable
side, we exclusively recommend one of the following two connectors.

Manufacturer:

Phoenix Contact GmbH & Co. KG
Flachsmarkt 8
32825 Blomberg
Germany
http://www.phoenixcontact.com

Types:

VS-09-PROFB-SC
(Phoenix article no. 1654549, with screwed contacts,
available from YSI under order number 902 888Y)

or:

VS-09-PROFB-SP
(Phoenix article no. 1654345, with spring contacts)

To remove the blind plug, lift the safety bracket with a suitable screw
driver at both clamps on the left and right side (see Fig. 7-1).