The antenna will typically be shipped with the Profiler T-frame assembly. The antenna cable
will already be connected to the Controller System, Port J5 as shown in Figure 3-5. Unpack and
reattach the antenna base to the top plate of the T-frame. Mounting holes and hardware are
included.

Figure 3-12 Remounting the Cellular Antenna during Equipment Staging


DEPTH SOUNDER (optional)

The Depth Sounder is a sensor that periodically confirms water depth as directed by the profiling
program downloaded to the Campbell CR10X data logger. The sensor unit, if ordered, will be
mounted to the platform deck and the cable will be plugged into Port J6 on the Controller
Assembly (PVC cylindrical enclosure). Check to make certain the sensor unit is securely
fastened and the cable is appropriately routed and connected.

METEOROLOGICAL SENSORS (optional)

Some or the entire MET suite of sensors may be mounted to your Profiler System. Check that
the sensors you ordered are present and mounted. Some sensor units will need to be remounted
if they were separately packed for shipping. The MET suite cabling will typically require a PVC
breakout box to connect cables from separate units. One cable then plugs into Port J7 (installed
as needed) of the Controller Assembly. The cabling should be in place. Insure secure mounting
of sensor units and proper routing and connection of all cables. Live testing will be described
prior to final deployment.

Unless you have additional options for your Profiler platform, such as beacons or other optional
equipment, the indoor staging procedure is complete. The next step will be to setup and connect
the CR10X in the Controller Assembly via direct and/or wireless connection using the
LoggerNet™ software.