YSI Profiler System USER Manual

In Step 5 (Sonde parameters) you must record in the box on the right all parameters you have
chosen to monitor. You probably have jotted these down in notes. If not, return to the sonde
menu and check the Report setup to make sure you have listed all parameters, in the same order
as in Step 5 of the setup wizard.. Depth in meters is a given, and in fact, is listed by default in
the Sonde report output column. You now click on each parameter and then click Add >> to add
it to the list on the right. You may do this one at a time, or you may press and hold the Control
key (Ctrl) on your keyboard and click on the appropriate parameters in the list on the left. The
Add button becomes active. Click Add >> and the list on the right appears as shown below for
this example. You may now need to rearrange the order of this list to match the sonde Report
setup.

It is essential that the list of parameters you create matches the parameters set in your sonde
(exact parameter, including units chosen and the order listed). For example, SpCond (mS/cm) is
not the same as SpCond (uS/cm) this does make a difference when the CR10X data logger and
sonde communicate during profiling routines. This makes a difference later on during post data
processing

The screen below shows an example list of parameters. There are nine (9) parameters total.
Depth in meters is a requirement. In Profile Wizard there is a maximum of twelve (12)
parameters that you may list, counting Depth in meters as one of the twelve.

IMPORTANT:
The order of parameters listed must match the Report setup in the sonde.


In the following figure Depth (meters) was moved so that it occurs after DO Chrg but before pH.
Parameters may be moved up and down the list by left clicking and dragging up or down.


YSI Environmental

669523 Rev B Page 2-15