Type 2-Stop sampling to discontinue sampling data, then type ‘0’ several times until you return
to Main menu, and then finally to the # sign. You have now exited the sonde menu and you may
now disconnect the sonde from the cable to the PC.

You have now completed checking the YSI sonde for proper configuration. In addition, you have
become generally familiar with all submenus in the sonde Main menu. If you have yet to
calibrate the sonde, refer to your sonde user manual and use the 2-Calibrate function to enter the
submenus to complete this operation.

Once calibration setup is complete ensure
the sonde has new batteries installed and the
sensor guard has the provided standard
weight attached. The top of the weight,
inside the sensor guard, should be painted
black to ensure proper operation of the
optical sensors. Refer to Figure 3.1.

Figure 3-1 Sonde Sensor Guard with

standard weight

NOTE:
Sonde must be equipped with provided sonde weight for the YSI Profile system to work
properly.

#

xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx

-----------------Main------------------
1-Run 5-System
2-Calibrate 6-Report
3-File 7-Sensor
4-Status 8-Advanced

===============================================================
Temp SpCond Cond Sal DO DOchrg Depth pH pH Turbid+ Chl Fluor Battery
C mS/cm mS/cm ppt mg/L meters mV NTU ug/L %FS volts
-----------------------------------------------------------------------------------------
*** 1-LOG last sample 2-LOG ON/OFF, 3-Clean optics ***
Stabilizing:4 3 2 1
16.93 0.001 0.001 0.00 8.35 53.2 0.000 1.77 300.2 428.2 0.2 0.0 12.3
16.92 0.001 0.001 0.00 8.35 53.1 0.000 2.04 284.9 428.2 0.2 0.0 12.3
16.91 0.001 0.001 0.00 8.36 53.2 0.000 2.19 276.4 428.2 0.2 0.0 12.3
16.91 0.001 0.001 0.00 8.36 53.1 0.000 2.36 266.9 428.1 0.2 0.0 12.3
16.91 0.001 0.001 0.00 8.36 53.2 0.000 2.47 260.7 428.1 0.2 0.0 12.3
16.91 0.001 0.001 0.00 8.36 53.2 0.000 2.56 255.6 428.1 0.2 0.0 12.3