YSI Profiler System USER Manual

------------Advanced setup-------------
1-(*)VT100 emulation
2-( )Power up to Menu
3-( )Power up to Run
4-( )Comma radix
5-(*)Auto sleep RS232
6-(*)Auto sleep SDI12
7-( )Multi SDI12
8-( )Full SDI12

Note that both RS232 and SDI12 Auto sleep functions are enabled. If not, change these settings
by typing the number of the function to toggle on or off the asterisk. Next, return to the
Advanced menu and then type 3-Sensor.

---------------Advanced----------------
1-Cal constants
2-Setup
3-Sensor
4-Data filter

Note in the Advanced sensor screen below that DO warm up time is set to 60 seconds, which is
the appropriate setting for the profiling routine. Since there are sensors with wiping mechanisms
in this sonde, keep in mind that the SDI12-M/wipe=1 means that the sensor will wipe before
each reading (turbidity or chlorophyll). If you set this number to two (2), a wipe will occur
before every other measurement. Choose the number appropriate for your water conditions.

------------Advanced sensor------------
1-TDS constant=0.65
2-Latitude=40
3-Altitude Ft=0
4-DO temp co %/C=1.1
5-DO warm up sec=60
6-( )Wait for DO
7-Wipes=1
8-Wipe interval=5
9-SDI12-M/wipe=1
A-Turb-6136 temp co %/C=0.6
B-(*)Turb-6136 spike filter
C-Chl temp co %/C=0

Next, make a quick check of other Advanced settings, specifically Data filter Cal and constants.
Type ‘0’ to return to Advanced.

---------------Advanced----------------
1-Cal constants
2-Setup
3-Sensor
4-Data filter

YSI Environmental

669523 Rev B

Page 3-5