YSI Profiler System USER Manual

Appendix C

I/O Module Configuration

The SDM SIO4 module needs configured before it can be used. Connect the I/O module to 12
VDC (either a battery or the 6960 main power). The STATUS indicator should turn on briefly after
power up. Connect a PC to the I/O module at Port 1. You will need a female/female null adapter.

Start Hyperterminal and then select the PC COM port. Set it to9600 baud. Press the SDMSIO reset
button, and the STATUS indicator should turn on. The unit will scroll SDMSIO4->. If these
letters do not appear, either the power is not on or there may be a faulty RS-232 connection to the
PC.

From the Hyperterminal menu select the Transfer button and then select SEND TEXT FILE.
Locate the SDM_info .txt file and send it. You should see the following:

…………………………………………..
portset 2 12 3 0 0 0 0
portset 3 11 3 0 0 0 0
portset 4 13 3 0 0 0 0
strst 1 " &0D&0A &0D&0A"
strst 3 "para&0D&0A"
strst 4 "exit&0D&0A"
strst 5 "data&0D&0A"
strst 9 "sleep&0D&0A"
strst 11 "run&0D&0A"
strst 12 "fltreset&0D&0A"
strst 15 "&1B&0D&0A"
strst 49 "&0D&0A&0D&0A&0D&0A"
strst 20 "*H&0D&0A"
strst 21 "*M&0D&0A"
strst 22 "*S&0D&0A"
strst 23 "*A&0D&0A"
strst 24 "*Q&0D&0A"
strst 25 "*C&0D&0A"
fltst 6 "fffffffffffffffffffffffff"
fltst 7 "zxt[$SDDBT,]t[,f,]FXz"
fltst 8 "xt[$SDDPT,]FXz"
fltst 10 "f"
fltst 50 "i[$]f"
fltst 51 "i[$]n6Fi[MfF]"
fltst 52 "F"
fltst 53 "FCF"
fltst 54 "zft[,]f"
fltst 55 "zfn1f"
fltst 56 "Ft[,]Ft[,]F"
fltst 57 "ft[,]ft[,]f"
exit
………………………………………………………

At this point press enter and then power cycle the Profiler unit. Press the reset button. The unit
should scroll the following header, SDMSIO4-> . Type Strrd 12 and press enter. You should see
the following response SDMSIO4-> fltreset .

Type exit escape (SDMSIO4->exit).