YSI Profiler System USER Manual

Since you depend on power to the sonde for proper calibration and measurements in the field,
we recommend this in the report choices.

-------------Report setup--------------
1-( )Date B-(*)DOchrg
2-( )Time hh:mm:ss C-( )Press psig
3-(*)Temp C D-(*)Depth meters
4-(*)SpCond mS/cm E-(*)pH
5-(*)Cond mS/cm F-(*)pH mV
6-( )Resist G-(*)Turbid+ NTU
7-( )TDS H-(*)Chl ug/L
8-(*)Sal ppt I-(*)Fluor %FS
9-( )DOsat % J-(*)Battery volts
A-(*)DO mg/L

Next, type ‘0’ to return to the Main menu, and then select 5-System to check on the SDI-12
address. The sonde SDI-12 address must be set to 0 in order to communicate with the Profiler.

-----------------Main------------------
1-Run 5-System
2-Calibrate 6-Report
3-File 7-Sensor
4-Status 8-Advanced

-------------System setup--------------
1-Date & time
2-Comm setup
3-Page length=25
4-Instrument ID=99F0932 AA
5-Circuit board SN:0001066E
6-GLP filename=0001066E
7-SDI-12 address=0
8-Language:English

While you are in System menu, check 2-Comm setup to make certain that the parameters Auto
baud and 9600 baud rate are enabled. Once you verify this, type ‘0’ two times to back out to the
Main menu.

--------------Comm setup---------------
1-(*)Auto baud 5-( )2400 baud
2-( )300 baud 6-( )4800 baud
3-( )600 baud 7-(*)9600 baud
4-( )1200 baud 8-( )19200 baud

YSI Environmental

669523 Rev B

Page 3-3