NOTE:
When you select a Flag or a Port, it typically takes affect immediately. However, the
response may not appear right away because of ‘table repetition rates’.

At this point Flag 2 will stay on indefinitely. We recommend that all pre-deployment tests be
performed before you start the profiler program. The procedure to start the profiler program will
be described again below.

After pre-deployment testing you may start the profiler program by turning Flag 2 OFF and then
turn Flag 5 ON. This will force a profile sequence within about 60 seconds (the repetition rate
of the display).

Although the above Flag and Port names can be renamed to be more meaningful to you, we
recommend that you learn the terminology at first before making these changes. See Appendix
D for Sonde Parameter ID numbers, which may of particular interest in depth profiling
applications.

RECOMMENDED PRE-DEPLOYMENT TESTS


NOTE:
Leave Flag 2 ON during all pre-deployment testing. This is setup mode. The system will
not enter a programmed profiling routine until you invoke the command 3.

Use the parameters and values below in the Numeric Display along with explanations of each
pre-deployment test.


3

If you are confident that all is working, you may start the profiler. To do this turn Flag 2 OFF and then

turn Flag 5 ON.