YSI Profiler System USER Manual

Section 5

Pre-Deployment Testing

5.1 Setting Up LoggerNet™ for System Testing

In this section you will use LoggerNet™ control of the CR10X profiler program to test various
components of the Profiler System prior to deployment activities. The following components
will be tested, unless the component is optional and not present.

Sonde response (RS-232 and SDI-12)

Winch motor control

Depth sounder or Fixed reference sonde

Compass

Meteorological sensors

Communications

Before proceeding you should review operation of the Campbell Scientific CR10X Data Logger
and LoggerNet™ application software. Review the LoggerNet™ Quick Start Guide, and if
necessary, the LoggerNet™ user’s manual.

The procedure in this section assumes that you have completed LoggerNet™ Station
configuration and setup. If not, then refer to information in previous sections of this manual. In
particular, verify that the YSI 6-series sonde is set up as specified in Section 3.2, Setting Sonde
Parameters. Verify that the sonde is

Not set to log in Unattended Mode

Not set to report date and time in Report setup

Is set to SDI-12 address = 0

Is set to report depth in meters

Refer to Section 3 of this manual if you need to verify any of the above sonde parameters.