2 testing – YSI Vertical Profiler Systems User Manual

Page 65

Advertising
background image

YSI Profiler System USER Manual

5.2 Testing


You are now ready to reconnect the CR10X and begin testing components of the Profiler
system. Do the following and verify that ports and flags behave as described below.

1. If necessary, power up the CR10X (Controller Assembly connector).

2. Wait about 10 seconds, and then press the Connect button.

3. If the CR10X was previously programmed, wait another 30 seconds for Flag 1 and Flag 2

to turn ON (green lights appear in icons).

4. Flag 1 will turn OFF after the CR10X initializes (a few seconds). At that moment you

should hear a relay click ON/OFF from within the Controller housing.

The Ports and Flags display should now have Flag 2 illuminated green to indicate that it is set
high or ON. This is Setup mode and the program remains idle until you prompt it.


NOTE:
If Flag 1 or Flag 2 does not turn ON, the program was not uploaded correctly.


NOTE:
If Flag 1 does not turn OFF after being on for a few seconds, the program was not
uploaded correctly.


After Flag 1 turns OFF, remain connected to verify that the numeric display starts to update.
Normally the updates occur every 10 to 20 seconds, starting approximately at the top of the
minute. (Installed Meteorological sensors update every 60 seconds.).

YSI Environmental

669523 Rev B

Page 5-6

Advertising