6 measuring with increment procedure (methods), 1 selecting the measuring method, Measuring with increment procedure (methods) 9.6.1 – Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual
Page 113: Selecting the measuring method

ProLab 4000
Ion concentration
113
ba75646e02
01/2010
9.6
Measuring with increment procedure (methods)
9.6.1
Selecting the measuring method
The following methods are supported:
z Standard addition
z Double standard addition
z Standard subtraction
z Sample addition
z Standard subtraction
z Blank value addition (Standard addition with blank value correction)
1
Perform the preparatory activities (see page 103).
2
Highlight the channel of the electrode in the measured value
display.
3
If necessary, select the ISE display with <MODE>.
4
Thoroughly rinse the electrode with distilled water.
5
Adjust the temperature of the calibration standard solutions.
6
Select a unit in the menu, Measuring / ISE settings / Units .
7
Select a method in the menu, Measuring / ISE methods .
The selected method is started (see page 114 ff).