Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual
Page 120

Ion concentration
ProLab 4000
120
ba75646e02
01/2010
6
In the Sample register, enter the values for the sample volume
(Sample volume) and the volume of the ISA/TISAB solution
(ISA/TISAB volume) as necessary.
7
Check the currently displayed ions.
If the ion of the standard (e.g. Cl (Sub) is displayed incorrectly:
Change the ion type of the standard solution [e.g. Cl (Sub)] in
the ION register.
8
Thoroughly rinse the electrode with deionized water.
9
Immerse the electrode in the sample.
10
Start the measurement with [Continue] .
The measured value is checked for stability (Stability control).
11
Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Standard register opens up.
12
In the Standard register, enter the values for Standard volume
and Standard concentration if necessary.
13
Add the standard solution to the sample.
14
Start the measurement with [Continue].
The measured value is checked for stability (Stability control).
15
Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Calculation register with the calculated measured value
pops up.