16 what to do if, 1 ph and orp measurement, Chapter 16 w – Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual
Page 189: To chapter 16 w

ProLab 4000
What to do if...
189
ba75646e02
01/2010
16
What to do if...
16.1
pH and ORP measurement
Error message Err1
Error message Err2
Error message Err4
Error message, Error
Cause
Remedy
pH electrode:
– Measured value outside the
measuring range
– Use a suitable electrode
– Air bubble in front of the dia-
phragm
– Remove air bubble
– Air in the diaphragm
– Extract air or moisten junc-
tion
– Cable broken
– Replace the electrode
– Gel electrolyte dried out
– Replace the electrode
Cause
Remedy
– No electrode connected
– Connect the electrode
– Setting time during calibration
too long
– Adjust temperature if neces-
sary
– Recalibrate
Cause
Remedy
– Temperature not stable during
calibration.
– Adjust temperature if neces-
sary
– Recalibrate
Cause
Remedy
pH electrode:
– The values determined for zero
point and slope of the electrode
are outside the allowed limits.
– Recalibrate
– Diaphragm contaminated
– Clean diaphragm