7 ph, 1 general information, General information – Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual
Page 69

ProLab 4000
pH
69
ba75646e02
01/2010
7
pH
7.1
General information
Temperature
measurement
For reproducible pH measurements, it is essential to measure the tem-
perature of the test sample.
You have the following options to measure the temperature:
z Automatic temperature measurement with
– Temperature sensor (NTC 30 or Pt 1000) integrated in electrode
(if available).
– the integrated temperature sensor of another electrode that is
connected and immersed in the sample at the same time.
– an external temperature sensor (NTC 30 or Pt 1000).
z Manual determination and input of the temperature.
The display of the temperature indicates the active temperature mea-
suring mode:
Temperature
sensor
Resolution of
the temp. dis-
play
Mode
yes
0.1 °C
Automatic with temperature
sensor
-
1 °C
Manual