Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual
Page 129
Advertising

ProLab 4000
Ion concentration
129
ba75646e02
01/2010
Measuring
6
In the Sample register, enter the following values:
z the volume of the sample (Sample volume)
z the volume of the ISA/TISAB solution (ISA/TISAB volume)
z the volume of the blank value solution (Blank value volume)
z the concentration of the blank value solution (Blank value
conc.).
7
Thoroughly rinse the electrode with deionized water.
8
Immerse the electrode in the sample that was supplemented
with blank value solution.
9
Start the measurement with [Continue] .
The measured value is checked for stability (Stability control).
10
Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Standard register opens up.