2 ion selective measurement – Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual
Page 191

ProLab 4000
What to do if...
191
ba75646e02
01/2010
CalClock on red
background
16.2
Ion selective measurement
Error message Err1
Obviously incorrect
measured values
Error message, Error
(invalid calibration)
– Measurement procedure not
suitable
– Follow special procedure
Cause
Remedy
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring
system
Cause
Remedy
– Measuring range exceeded
– Dilute test sample
Cause
Remedy
– Electrode not connected
– Connect the electrode
– Cable broken
– Replace the electrode
Cause
Remedy
ISE electrode:
– Moisture in the plug
– Dry plug
– Electrode obsolete
– Replace the electrode
– Electrode unsuitable for the
range to be measured
– Use a suitable electrode
– Socket damp
– Dry socket
Calibration procedure:
– Wrong sequence of standards
for three point calibration
– Select correct sequence
– Calibration standards do not
have the correct temperature
(max.
±
2 °C temperature differ-
ence)
– Adjust the temperature of
the calibration standards