3 measurement settings – Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual
Page 134

Conductivity
ProLab 4000
134
ba75646d02
01/2010
10.2.3
Measurement settings
The following measurement settings are possible for conductivity mea-
suring cells:
Menu item
Setting
Description
Measuring
– Cell constant
– Calibrated cell constant
Measuring cells whose cell constant
is determined by calibration in KCL
control standard (calibration ranges
see page 149).
The currently valid cell constant is
displayed in the status line.
– Adjustable cell constant...
Any measuring cells with freely
adjustable cell constants in the range
0.250 to 25.000 cm
-1
.
– Adjustable LF 313T...
Measuring cell LF313T(-ID), nominal
cell constant 0.100 cm
-1
.
The cell constant can be adjusted in
the range 0.090 to 0.110 cm
-1
.
– LF 213T
Measuring cell with a nominal cell
constant of 0.010 cm
-1
.
The cell constant is permanently set.
– Temperature coefficient
– TC nLF
Nonlinear temperature compensa-
tion
– TC Lin...
0.000 ... 3.000 %/K
Coefficient of the linear temperature
compensation.
– TC off
No temperature compensation.
– TC nLin1......TC nLin4...
Measurement with entered or deter-
mined temperature coefficient.