Ion concentration
ProLab 4000
118
ba75646e02
01/2010
10
Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Standard register opens up.
11
Add standard solution to the sample (2 percent by volume of
the sample solution).
12
Start the measurement with [Continue].
The measured value is checked for stability (Stability control).
13
Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Calculation register with the calculated measured value
pops up.
14
Change the displayed unit of the measured value with [Unit] if
necessary.
15
If required, display the record of the current measurement with
[Record].
16
Start measuring further samples with [Continue].
Repeat the steps 2 - 14 for all samples.
or
Terminate the measuring method with [Cancel] and confirm the
safety query with [OK].
Measurement with the selected method is completed.