Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual
Page 137

ProLab 4000
Conductivity
137
ba75646d02
01/2010
Temperature compensa-
tion for diluted solutions
For the measurement of aqueous solutions
z in a low concentration range (diluted solution)
z in a small temperature range (|T - Tref| < 10 K)
z with a fixed reference temperature at 20 °C or 25 °C,
the procedures of linear (TC lin) and nonlinear (TC nLF) temperature
compensation are suitable.
The calculation of the temperature compensation is based on the pre-
set reference temperature, 20 °C or 25 °C. It appears on the display as
[TR20] or [TR25].
Application tips
Select the following temperature compensation given in the table
according to the respective test sample:
Note
The reference temperature and temperature compensation are set in
the Measuring menu (see page 134).
Test sample
Temperature compensation
Display
indica-
tor
Natural water
(ground water, surface
water, drinking water)
nLF
according to EN 27 888
nLF
Ultrapure water
nLF
according to EN 27 888
nLF
Seawater
(salinity)
Automatic nLF according to IOT
Sal, nLF
Other diluted solutions
with known tempera-
ture dependency
Linear temperature compensa-
tion;
Temperature coefficient can be
set in the range
0.001 ... 3.000 %/K
lin
Solutions with
unknown temperature
dependency
No temperature compensation