Installation

24

3.5.2.1 si792(x) D (4–20 mA/HART) wiring

DANGER
Explosion hazard. Never connect items to the transmitter that are
not specified on the control drawing/ATEX EC-Type certificate.
Refer to

Figure 6

and

Table 2

and the following steps to connect the

power supply to the si792(x) D transmitter.

Figure 4 si792(x) D (4–20 mA/HART) wiring

1 HART connection (see warnings in

section 3.6 on page 30

)

2 Wiring terminals—see

Table 4

Table 2 Terminal assignments—si792(x) D

Terminal No. Assignment

9

Equipotential bonding for ATEX. Do not connect for FM and CSA.

10

4–20 mA output (–)

11

4–20 mA output (+)