Calibration

76

4. Press

CAL

, enter passcode: 1105,

ENTER

to access the product

calibration once more. The display will briefly show CAL PRD and
then CALC.

5. Edit the displayed value to match the value measured by the

laboratory or portable meter and press

ENTER

. The new slope and

zero point will be shown.

6. Press

ENTER

to end the calibration. Return the sensor to the

process. The measured value will be shown in the display
alternately with Hold.

7. Press

ENTER

to return to the measure mode.

Note: For comparison calibration via Foundation Fieldbus, see

section 6.2.4 on page 63

.

8.2 Temperature sensor adjustment

Complete the following steps to calibrate the temperature sensor.
1. Press

CAL

, enter passcode: 1015,

ENTER

. CAL TMP will be

displayed for 3 seconds. The process temperature will be shown.

2. Measure the process temperature with an external thermometer.
3. Use the

UP ARROW

and

RIGHT ARROW

to enter the value of the

temperature from the external thermometer.

4. Press

ENTER

to save the value.

5. The transmitter remains in HoLD mode

.

Press

ENTER

again.

The transmitter remains in HoLD mode for approximately
20 seconds after calibration and returns to the measure mode.

Note: Repeat input of the process temperature value when “Err”
is displayed.