Operation—si792(x) D (4–20 mA/HART)

40

5.2.2 Temperature sensor configuration

An overview of the temperature compensation setup menu is shown in

Table 10

.

Press

ENTER

to access a menu item. Use the

ARROW KEYS

to edit

values. Press

ENTER

to save the settings. If a value is outside of the

acceptable range, “Err” will be displayed and the value will not be
accepted. To exit the menu and return to the measurement mode, press

CONF

and then

ENTER

.

5.2.3 Correction menu configuration

An overview of the corrections menu is shown in

Table 11

.

Table 10 Temperature sensor setup menu—tc

MNU

Select temperature unit

tc.UnIT

°C (default)
°F

Select temperature sensor

tc.rTD

22NTC (default)
30NTC

Table 11 Corrections setup menu—Corr

MNU

Specify the polarization voltage
Co.UPoL

Enter the polarization voltage (400–1000 mV; default: 675 mV)

Select pressure units

Co.UniT

Bar (default)
KPa
PSI

Specify the correction for process pressure (for % saturation only)
Co.PrES

Enter the process pressure (default: 1.013 Bar)

Specify the correction for salinity
Co. SAL

Enter the salinity value of the sample (0–45.00 ppt

1

; default: 0 ppt)

1

ppt—parts per thousand (corresponds to g/kg)