93

Appendix A Passcode editor

For applications requiring compliance with FDA 21 CFR Part 11, certain
device functions must be protected with editable passcodes. If
passcode protection is not required, use the preset passcodes
(

Table 27

).

To activate the passcode editor:
1. Press CONF.
2. Enter 1989, the administrator passcode. The initial screen will show

for approximately 3 seconds, then proceed to the next function.

3. Use the ARROW keys to change the passcode.
4. Press ENTER to proceed to the next function. Press CONF to exit

the passcode editor. Refer to

Table 27

for default passcodes.

Table 27 Default passcode settings

Display

Function

Default setting

Cal INFO

Calibration information

0000

CAL 0-CAL

Calibration zero

1001

CAL CAL

Calibration Saturation/Concentration

1100

CAL PROD

Product calibration

1105

CAL RTD

Temperature sensor adjustment

1015

CFG ERR

Error information

0000

CFG CONF

Configuration

1200

CFG SNSR MO Probe monitor

2222

CFG OUT

Current source

5555

CFG SPCL ST

Administrator passcode. Use

ARROW

keys to

select NO or YES

1989

NO SPCL ST

Do not change
administrator passcode

Press

ENTER

to return to the default

passcode
Press

CONF

to exit the editor

YES SPCL ST

Change administrator
passcode

Press

ENTER

to accept the new

passcode
Press

CONF

to exit the editor

Advertising