1 output configuration, Section 5.2.1 – Hach-Lange SI792 D, SI792x D, SI792x D-FF, SI792x D-PA User Manual User Manual

Page 38

Advertising
background image

Operation—si792(x) D (4–20 mA/HART)

36

To exit the configuration mode at any time, press

CONF

and then

ENTER

. The output current will be held for 20 seconds and the

measured value will be displayed.

Note: During configuration the transmitter remains in the Hold mode for
safety reasons. The loop current is frozen at the value specified in the
o1.HoLD menu option. The Sensoface icon is inactive. The
configuration mode indicator is displayed (

Figure 12 on page 33

).

5.2.1 Output configuration

o1

4–20 mA current output

1200

tc

Temperature sensor

Corr

Correction for pressure,
salinity, polarization voltage.

CAL

Calibration setup

AL

Alarm settings

Table 9 Output setup menu—out.1

MNU

Select the parameter

o1. FCT

DO (concentration by weight; default)
GAS (concentration by volume)

Select the sensor type

o1.SnSR

Type A (standard; default) examples: Oxysens, Oxyferm
Type B (trace measurements) example: Oxygold G

Select the units

o1.UniT

DO (default)

SAt % (percent saturation; default)
Conc µg/L
Conc mg/L
Conc ppb
Conc ppm

GAS

GAS % (default)
GAS ppm

Table 8 Configuration menu