2 do calibration (default), 1 calibration in air (recommended), Calibration – Hach-Lange SI792 D, SI792x D, SI792x D-FF, SI792x D-PA User Manual User Manual

Page 75

Advertising
background image

Calibration

73

8.1.2

DO calibration (default)

Make sure that DO is selected in the configuration menu (

Table 9 on

page 36

). For a 2-point calibration, calibrate the zero point first

(

section 8.1.1 on page 72

).

8.1.2.1 Calibration in air (recommended)
1.
Press

CAL

, enter passcode: 1100,

ENTER

. The display will briefly

show CAL and then the entered (or default) humidity value.

2. Remove the sensor from the process and place in air.
3. Enter the actual relative humidity and press

ENTER

. The display will

show the entered (or default) barometric pressure.

4. Enter the barometric pressure and press

ENTER

. A drift check will

begin. The drift check may take some time. When complete, the
display will show the concentration in the units specified in the
configuration menu.
Note: To exit the drift check after, 10 seconds, press

CAL

(accuracy

will be reduced).

5. Accept the calculated value for the concentration or adjust the value

as necessary and then press

ENTER

. The display will show the new

slope and zero (relative to 25 °C and 1.013 mbar).

6. Press

ENTER

to end the calibration. Return the sensor to the

process. The measured value will be shown in the display
alternately with Hold.

7. Press

ENTER

to return to the measure mode.