2 profibus pa communication, Section 7.2 on, Operation—profibus pa – Hach-Lange SI792 D, SI792x D, SI792x D-FF, SI792x D-PA User Manual User Manual

Page 69: 2 configuration menu (continued)

Advertising
background image

Operation—Profibus PA

67

7.2 Profibus PA communication

Profibus PA uses a master/slave data exchange technique. The master
(typically a PLC) generates queries to individual slaves. The slaves, in
turn, reply back with a response to the master. A Profibus message
contains the information required to send a query or request, including
the slave address, function code, data, and a checksum. See

Table 20

on page 67

for Profibus communication parameters.

Enter the salinity

0–45.00 SAL (default: 00.00)

Enter polarization voltage
U-Pol

0 mV to 800 mV (default: 675 mV, 0mV = Off)

Select Sensocheck

CHECK ON

Continuous Sensocheck
evaluation of sensor function

CHECK OFF (default)

Select the calibration mode

CAL

Saturation (Sat)
Concentration (Conc) (default)

Enter cal timer interval

0000–9999 h (default: 0000 h (Off))

Enter Profibus address
Edit

0001–0126 (default: 0126)

Table 20 PROFIBUS PA communication

Cause

No. of

binary

message

Analog input
status

Physical Block
(PB)
Global status

Text of binary m
(default)

Factory settings
defective

1

0000 11xx

Failure

ERR SYSTEM

Configuration data
defective, Gaincheck

2

0000 11xx

Failure

ERR PARAMET

7.1.2

Configuration menu (continued)