4 calibration errors, Troubleshooting – Hach-Lange SI792 D, SI792x D, SI792x D-FF, SI792x D-PA User Manual User Manual

Page 84

Advertising
background image

Troubleshooting

82

10.4 Calibration errors

Table 25

describes the calibration error messages.

Table 25 Calibration error messages

Display

Description

Corrective action

Slope error icon

Slope is out of range

Make sure that the specified
values in the configuration
menu are correct.
Make sure of the
temperature specification in
configuration.
Replace the sensor.

CAL ERR with
calibration timer
icon

Calibration canceled due to
large drift in mV reading.

Make sure that the
temperature does not
change during calibration.
Make sure that the cable is
sufficiently shielded from
strong electric fields.
Make sure that the sensor
has electrolyte.
Clean the sensor.
Replace the sensor.

Advertising