2 standard analog input block (ai), 1 operating modes, 2 set the parameter and units – Hach-Lange SI792 D, SI792x D, SI792x D-FF, SI792x D-PA User Manual User Manual
Page 51: Operation—foundation fieldbus

Operation—Foundation Fieldbus
49
6.2.2 Standard analog input block (AI)
Three Analog Input Function Blocks provide for cyclic transmission of
measured values (currently measured value with status, alarm limits,
freely selectable process parameter).
6.2.2.1 Operating modes
Use the MODE_BLK parameter to set the following operating modes:
•
OOS—out of service. If not write-protected, access to all
parameters is allowed.
•
MAN—manual
•
Auto—online, normal state
6.2.2.2 Set the parameter and units
Use CHANNEL to set the measured parameter and units (
The corresponding measurement unit is selected in the UNITS
subparameter of XD_SCALE (
Table 16 on page 51
).
Table 15 Measurement parameters and units
Channel
Parameter
Unit
Unit value
1
Saturation %
(percent)
1342
2
Concentration
mg/l
2001
μg/l
2002
ppm
1423
ppb
1424
3
Temperature
°C
1001
°F
1002
4
Zero
A
2003
5
Slope
A
1587
6
Volume
concentration
% percent
1342
ppm
1423