37

Press

ENTER

to access a menu item. Use the

ARROW KEYS

to edit

values. Press

ENTER

to save the settings. If a value is outside of the

acceptable range, “Err” will be displayed and the value will not be
accepted. To exit the menu and return to the measurement mode, press

CONF

and then

ENTER

.

Example: Set the 4 mA current output signal to 2 mg/L O

2

.

1. Press

CONF

, enter passcode: 1200,

ENTER

. The display will show

Conf and then out.1MNU.

2. Press

ENTER

to access the output setup menu. The display will

show o1.UniT.

3. Press

ENTER

,

ENTER

to show the 4 mA submenu. The display will

show o1.4mA.

4. Use the

UP ARROW

and

RIGHT ARROW

to edit the value to

03.00 mg/L. Press

ENTER

to save the value. The display will show

o1.20mA.

5. Press

CONF

and then

ENTER

to exit the configuration menu.

Specify the value for the 4 mA signal (

section 5.2.1.1

)

o1. 4mA

Enter the minimum DO or GAS value (default: 0%)

Specify the value for the 20 mA signal (

section 5.2.1.1

)

o1.20mA

Enter the maximum DO or GAS value (default: 500%)

Set time averaging filter for reducing noise (

section 5.2.1.2

)

o1.Ftme

Enter the time in seconds (0 to 120 seconds; default: 0 seconds)

Send a 22 mA signal during errors (

section 5.2.1.3

)

o1.FAil

On
Off (default)

Specify the value to send during HOLD periods (

section 5.2.1.4

)

o1.HoLD

LAST measured value (default)
FIXed

Enter the value in mA (0 to 22 mA; default: 21 mA)