Hach-Lange EVITA OXY User Manual User Manual
Page 21
Advertising
19
1. Place the transmitter in free air so that the OXY
1100 sensor points downwards, see fi g. 30.
2. Turn on the power to the signal converter USC
5000/6000/7000.
3. Leave the transmitter in this position for at least
1 hour to allow the sensor to stabilise before
calibrating.
Fig. 30
4. Check factory settings, see p. 21. Change if
necessary.
5. Reset the lifetime counter, see fi g. 32, p. 22.
6. Calibrate in accordance with the instructions in
the “Calibration” section, p. 31.
7. After calibration is completed, place the
transmitter in the measuring media.
System with USC 5000/6000/7000
Advertising