Symptom

Current output

Event code

Cause

Remedy

None

Undefi ned

IU00

Power on

Registration of time of power on

IU01

Add-on module applied

Registration of applied add-on module

IU02

Add-on module defective or removed

Replace or install add-on module

IU03

Parameter successfully corrected

Information

Empty display

1. Supply voltage
2. USC 5000/6000/7000 defective

1. Check supply voltage
2. Replace USC 5000/6000/7000

Dots in upper line of display

(I out 1)

EU61

See EU61 p. 45

Same as EU61 p. 45

(I out 2)

EU62

See EU62 p. 45

(I out 3)

EU63

See EU63 p. 45

(I out 4)

EU64

See EU64 p. 45

Flashing bell in display

Undefi ned

WU10

I out 1: Value at 4 mA >= Value at 20 mA

Check I out 1 setup

WU11

I out 2: Value at 4 mA >= Value at 20 mA

Check I out 2 setup

WU12

I out 3: Value at 4 mA >= Value at 20 mA

Check I out 3 setup

WU13

I out 4: Value at 4 mA >= Value at 20 mA

Check I out 4 setup

WU14

The measured value has exceeded the current output 1 setting

Change value at 20 mA on I out 1

WU15

The measured value has exceeded the current output 2 setting

Change value at 20 mA on I out 2

WU16

The measured value has exceeded the current output 3 setting

Change value at 20 mA on I out 3

WU17

The measured value has exceeded the current output 4 setting

Change value at 20 mA on I out 4

WU18

The measured value is below the current output 1 setting

Change value at 4 mA on I out 1

WU19

The measured value is below the current output 2 setting

Change value at 4 mA on I out 2

WU20

The measured value is below the current output 3 setting

Change value at 4 mA on I out 3

WU21

The measured value is below the current output 4 setting

Change value at 4 mA on I out 4