Symptom
Current output
Event code
Cause
Remedy
None
Undefi ned
IU00
Power on
Registration of time of power on
IU01
Add-on module applied
Registration of applied add-on module
IU02
Add-on module defective or removed
Replace or install add-on module
IU03
Parameter successfully corrected
Information
Empty display
1. Supply voltage
2. USC 5000/6000/7000 defective
1. Check supply voltage
2. Replace USC 5000/6000/7000
Dots in upper line of display
(I out 1)
EU61
See EU61 p. 45
Same as EU61 p. 45
(I out 2)
EU62
See EU62 p. 45
(I out 3)
EU63
See EU63 p. 45
(I out 4)
EU64
See EU64 p. 45
Flashing bell in display
Undefi ned
WU10
I out 1: Value at 4 mA >= Value at 20 mA
Check I out 1 setup
WU11
I out 2: Value at 4 mA >= Value at 20 mA
Check I out 2 setup
WU12
I out 3: Value at 4 mA >= Value at 20 mA
Check I out 3 setup
WU13
I out 4: Value at 4 mA >= Value at 20 mA
Check I out 4 setup
WU14
The measured value has exceeded the current output 1 setting
Change value at 20 mA on I out 1
WU15
The measured value has exceeded the current output 2 setting
Change value at 20 mA on I out 2
WU16
The measured value has exceeded the current output 3 setting
Change value at 20 mA on I out 3
WU17
The measured value has exceeded the current output 4 setting
Change value at 20 mA on I out 4
WU18
The measured value is below the current output 1 setting
Change value at 4 mA on I out 1
WU19
The measured value is below the current output 2 setting
Change value at 4 mA on I out 2
WU20
The measured value is below the current output 3 setting
Change value at 4 mA on I out 3
WU21
The measured value is below the current output 4 setting
Change value at 4 mA on I out 4
USC