Usc 6000 relay output settings, Fig. 38 – Hach-Lange EVITA OXY User Manual User Manual
Page 31

29
Output
USC
I out 1
USC
Timer interval
000. h
RELAY 1
1-999 hours
Active time
00 min
RELAY 1
1-59 min
Relay at
active Open
RELAY 1
Open/Closed
Relay at
warning Open
Open/Closed
RELAY 1
Relay at
error Open
RELAY 1
Open/Closed
RELAY 1
Relay 1
USC
Relay 2
USC
Relay 3
USC
~~
081R9414.02.02
Function
Off
RELAY 1
On
Timer
Off
Function
Error
RELAY 1
Error
Warning
Hi Limit
Lo Limit
Cleaning
Oxygen
Temperature
Value
Oxygen
Setpoint
050.00 mg/l
RELAY 1
0 - 999.99
Hysteresis
005.00 mg/l
RELAY 1
0 - 999.99
Setpoint
035.00 ºC
RELAY 1
Hysteresis
005.00 ºC
RELAY 1
Relay above
limit Open
RELAY 1
0 - 999.99
0 - 999.99
Open/Closed
Relay below
limit Open
RELAY 1
Open/Closed
*
*
*
*
Cleaning interval
000. h
RELAY 1
Cleaning time
00 min
RELAY 1
Relay at
cleaning Open
RELAY 1
1-999 h
1-59 min
Open/Closed
**
USC 6000 relay output settings
Setting of relay outputs 1 to 3 on USC 6000.
*
Units selected in “Transmitter basic settings” fi g. 35, p. 26
**
To allow the current output to stabilise after a cleaning, the
USC locks the current output during “Cleaning time” and
additional 10 minutes.
Same as Relay 1
Fig. 38
Same as Relay 1
Same as Hi Limit