Usc 7000 relay output settings, Fig. 39 – Hach-Lange EVITA OXY User Manual User Manual

Page 32

Advertising
background image

30

Output

USC

I out 1

USC

Timer interval

000. h

RELAY 1

1-999 hours

Active time

00 min

RELAY 1

1-59 min

Relay at

active Open

RELAY 1

Open/Closed

Relay at

warning Open

Open/Closed

Function

Error

RELAY 1

RELAY 1

Relay at

error Open

RELAY 1

Open/Closed

RELAY 1

Setpoint

050.00 mg/l

RELAY 1

0 - 999.99

Hysteresis

005.00 mg/l

RELAY 1

0 - 999.99

Relay above

limit Open

RELAY 1

Open/Closed

Relay 1

USC

Relay 2

USC

Relay 3

USC

~~

Relay below

limit Open

RELAY 1

Open/Closed

081R9415.02.02

Relay 4

USC

Relay 5

USC

Function

Off

RELAY 1

One transmitter

All Errors

All Warnings

Timer

Off

Relay at

warning Open

Open/Closed

RELAY 1

Relay at

Error Open

RELAY 1

Open/Closed

Select

transmitter no. 00

RELAY 1

00-15

Error

Warning

Hi Limit

Lo Limit

Cleaning

Oxygen

Temperature

Value

Oxygen

Setpoint

035.00 ºC

RELAY 1

Hysteresis

005.00 ºC

RELAY 1

0 - 999.99

0 - 999.99

*

*

*

*

Cleaning interval

000. h

RELAY 1

Cleaning time

00 min

RELAY 1

Relay at

cleaning Open

RELAY 1

1-999 h

1-59 min

Open/Closed

**

USC 7000 relay output settings

Setting of relay outputs 1 to 5 on USC 7000

*

Units selected in “Transmitter basic setting”, fi g. 35, p. 26.

**

To allow the current output to stabilise after a cleaning, the
USC locks the current output during “Cleaning time” and
additional 10 minutes.

Fig. 39

Same as Relay 1

Same as Relay 1

Same as Relay 1

Same as Relay 1

Same as Hi Limit