Hach-Lange EVITA OXY User Manual User Manual

Page 47

Advertising
background image

45

Symptom

Current output

Event code

Cause

Remedy

Flashing bell in display

Undefi ned

WU39

HART communication error

Check cables

WU40

TMS communication error

Replace TMS

WU41

A parameter is out of range.
It could not be replaced by its default value

Fixed at power-up

WU43

Too many errors at the same time.
Some errors are not registered correctly

Power-up USC

WU49

Internal error

Power-up USC

One or more
current outputs
= 3.8 mA or 24 mA

EU60

CAN bus application does not respond.
This error is shown when the maximum number of
communication attempts is exceeded.

1. Power-up USC
2. Replace USC

Undefi ned

EU61

USC unable to read data required for I OUT 1
1. No reply from transmitter
2. Transmitter has got new address
3. Another master confl icts on HART

®

comm.

4. Supply voltage for transmitter too low

5. Too much noise on current loop
6. Transmitter defective
7. USC defective

1. Check shielding and cable connections
2. Power-up USC
3. Disconnect the other master or set USC to

another master, see fi g. 51, p. 52

4. Check supply voltage at transmitter,

supply voltage must be at least 20 V

5. Check shielding and cable connections
6. Replace transmitter
7. Replace USC

EU62

USC unable to read data required for I OUT 2
1 - 7: Same as EU61

Same as EU61

EU63

USC unable to read data required for I OUT 3
1 - 7: Same as EU61

Same as EU61

EU64

USC unable to read data required for I OUT 4
1 - 7: Same as EU61

Same as EU61