Hach-Lange EVITA OXY User Manual User Manual

Page 52

Advertising
background image

50

USC

Type

USC 5000

USC 6000

USC 7000


No. of transmitters 1 (point-to-point installation)

1 (point-to-point installation)

15 (multidrop)


Measuring

Dissolved oxygen:

0 to 10 – 500 %

Dissolved oxygen:

0 to 10 – 500 %

Dissolved oxygen:

0 to 10 – 500 %

range

0 to 0.1 – 50mg/l or ppm

0 to 0.1 – 50mg/l or ppm

0 to 0.1 – 50mg/l or ppm

Temperature:

0 – 70°C (32°F – 158°F)

Temperature:

0 – 70°C (32°F – 158°F)

Temperature:

0 – 70°C (32°F – 158°F)


Measuring

Oxygen:

± 0.5 %

Oxygen:

± 0.5 %

Oxygen:

± 0.5 %

uncertainty

Temperature:

± 0.5°C (0.9°F)

Temperature:

± 0.5°C (0.9°F)

Temperature:

± 0.5°C (0.9°F)

Current outputs

One 4-20 mA (scaleable),

Two 4-20 mA (scaleable),

Four 4-20 mA (scaleable), isolated

24V active galvanic isolated

24V active galvanic isolated

24V active galvanic

Max. load: 800 ohm

Max. load: 800 ohm

Max. load: 800 ohm

Relay outputs

None

3 relay outputs SPST (max. 48 Va.c. or 30 Vrms 4 A)

5 relay outputs SPST (max. 48 Va.c. or 30 Vrms 4 A)

Digital input

Min. 12 to max 30 V d.c. (6 KOhm)

Min. 12 to max 30 V d.c. (6 KOhm)

Min. 12 to max 30 V d.c. (6 KOhm)

Display

Back-lit alphanumeric LCD display

Back-lit alphanumeric LCD display

Back-lit alphanumeric LCD display


Enclosure

IP 67 according to IEC 529; NEMA 4X

IP 67, IP 20 according to IEC 529; NEMA 4X

IP 67, IP 20 according to IEC 529; NEMA 4X

Ambient

Storage:

-40°C to +70°C (-40°F to 158°F)

Storage: -40°C to +70°C (-40°F to 158°F)

Storage: -40°C to +70°C (-40°F to 158°F)

temperature

Operation:

-40°C to +60°C (-40°F to 140°F)

Operation:-40°C to +60°C (-40°F to 140°F)

Operation: -40°C to +50°C (-40°F to 122°F)

Environment

Indoor use, altitude up to 2000 m

Indoor use, altitude up to 2000 m

Indoor use, altitude up to 2000 m

Pollution Degree II

Pollution Degree II

Pollution Degree II

Relative humidity 80% up to 31°C

(87.8°F)

Relative humidity 80% up to 31°C

(87.8°F)

Relative humidity 80% up to 31°C

(87.8°F)

decreasing linearly to 50% at 40°C (104°F)

decreasing linearly to 50% at 40°C (104°F)

decreasing linearly to 50% at 40°C (104°F)

Power supply

100-240 Vac +10%/-15%

100-240 Vac +10%/-15%

100-240 Vac +10%/-15%

50/60 Hz. 5-20VA

50/60 Hz. 5-20VA

50/60 Hz. 5-20VA

Fuse: T500mA/250V (Not to be changed by user)

Fuse: T500mA/250V (Not to be changed by user)

Fuse: T500mA/250V (Not to be changed by user)

11-30 Va.c./d.c. 5-20 W

11-30 Va.c./d.c. 5-20 W

Automatic

Compensating for temperature, barometric

Compensating for temperature, barometric

Compensating for temperature, barometric

calibration

pressure, humidity and salinity

pressure, humidity and salinity

pressure, humidity and salinity

Cable glands

Jacob GmbH, cat. no. 50.013 PA (PG13.5)

Jacob GmbH, cat. no. 50.013 PA (PG13.5)

Jacob GmbH, cat. no. 50.013 PA (PG13.5)

6-12 mm cable diameter

6-12 mm cable diameter

6-12 mm cable diameter

Blind plugs

Jacob GmbH, cat. no. 1013 PA

Jacob GmbH, cat. no. 1013 PA

Jacob GmbH, cat. no. 1013 PA

Approvals

CE and C-tick approved

CE and C-tick approved

CE and C-tick approved

Emission: EN 50081

Emission: EN 50081

Emission: EN 50081

Immunity: EN 50082

Immunity: EN 50082

Immunity: EN 50082

Advertising