Hach-Lange EVITA OXY User Manual User Manual
Page 52

50
USC
Type
USC 5000
USC 6000
USC 7000
No. of transmitters 1 (point-to-point installation)
1 (point-to-point installation)
15 (multidrop)
Measuring
Dissolved oxygen:
0 to 10 – 500 %
Dissolved oxygen:
0 to 10 – 500 %
Dissolved oxygen:
0 to 10 – 500 %
range
0 to 0.1 – 50mg/l or ppm
0 to 0.1 – 50mg/l or ppm
0 to 0.1 – 50mg/l or ppm
Temperature:
0 – 70°C (32°F – 158°F)
Temperature:
0 – 70°C (32°F – 158°F)
Temperature:
0 – 70°C (32°F – 158°F)
Measuring
Oxygen:
± 0.5 %
Oxygen:
± 0.5 %
Oxygen:
± 0.5 %
uncertainty
Temperature:
± 0.5°C (0.9°F)
Temperature:
± 0.5°C (0.9°F)
Temperature:
± 0.5°C (0.9°F)
Current outputs
One 4-20 mA (scaleable),
Two 4-20 mA (scaleable),
Four 4-20 mA (scaleable), isolated
24V active galvanic isolated
24V active galvanic isolated
24V active galvanic
Max. load: 800 ohm
Max. load: 800 ohm
Max. load: 800 ohm
Relay outputs
None
3 relay outputs SPST (max. 48 Va.c. or 30 Vrms 4 A)
5 relay outputs SPST (max. 48 Va.c. or 30 Vrms 4 A)
Digital input
Min. 12 to max 30 V d.c. (6 KOhm)
Min. 12 to max 30 V d.c. (6 KOhm)
Min. 12 to max 30 V d.c. (6 KOhm)
Display
Back-lit alphanumeric LCD display
Back-lit alphanumeric LCD display
Back-lit alphanumeric LCD display
Enclosure
IP 67 according to IEC 529; NEMA 4X
IP 67, IP 20 according to IEC 529; NEMA 4X
IP 67, IP 20 according to IEC 529; NEMA 4X
Ambient
Storage:
-40°C to +70°C (-40°F to 158°F)
Storage: -40°C to +70°C (-40°F to 158°F)
Storage: -40°C to +70°C (-40°F to 158°F)
temperature
Operation:
-40°C to +60°C (-40°F to 140°F)
Operation:-40°C to +60°C (-40°F to 140°F)
Operation: -40°C to +50°C (-40°F to 122°F)
Environment
Indoor use, altitude up to 2000 m
Indoor use, altitude up to 2000 m
Indoor use, altitude up to 2000 m
Pollution Degree II
Pollution Degree II
Pollution Degree II
Relative humidity 80% up to 31°C
(87.8°F)
Relative humidity 80% up to 31°C
(87.8°F)
Relative humidity 80% up to 31°C
(87.8°F)
decreasing linearly to 50% at 40°C (104°F)
decreasing linearly to 50% at 40°C (104°F)
decreasing linearly to 50% at 40°C (104°F)
Power supply
100-240 Vac +10%/-15%
100-240 Vac +10%/-15%
100-240 Vac +10%/-15%
50/60 Hz. 5-20VA
50/60 Hz. 5-20VA
50/60 Hz. 5-20VA
Fuse: T500mA/250V (Not to be changed by user)
Fuse: T500mA/250V (Not to be changed by user)
Fuse: T500mA/250V (Not to be changed by user)
11-30 Va.c./d.c. 5-20 W
11-30 Va.c./d.c. 5-20 W
Automatic
Compensating for temperature, barometric
Compensating for temperature, barometric
Compensating for temperature, barometric
calibration
pressure, humidity and salinity
pressure, humidity and salinity
pressure, humidity and salinity
Cable glands
Jacob GmbH, cat. no. 50.013 PA (PG13.5)
Jacob GmbH, cat. no. 50.013 PA (PG13.5)
Jacob GmbH, cat. no. 50.013 PA (PG13.5)
6-12 mm cable diameter
6-12 mm cable diameter
6-12 mm cable diameter
Blind plugs
Jacob GmbH, cat. no. 1013 PA
Jacob GmbH, cat. no. 1013 PA
Jacob GmbH, cat. no. 1013 PA
Approvals
CE and C-tick approved
CE and C-tick approved
CE and C-tick approved
Emission: EN 50081
Emission: EN 50081
Emission: EN 50081
Immunity: EN 50082
Immunity: EN 50082
Immunity: EN 50082