Symptom

Current output
signal

Event
code

Cause

Remedy

Error in measuring
value

Variable

Calibration value incorrect

1. Check FSO in atmospheric air. Hold the transmitter up

in the air with the sensor downwards. Check the value
against DS/EN 25814 (see appendix IIa)

2. Check zero point (see appendix IIb)
3. In the case of large reading error, recalibrate

Display switching
between “Oxygen” and
“End of calibration”

End of calibration phase

Go to “Send command” menu and start measuring, see fi g.
47, p. 36

Flashing bell in display

Depending on
the settings

WO10

The measured value has exceeded the transmitter output setting

Change upper range in “Basic settings”, see fi g. 35, p. 26

WO11

The measured value is below the transmitter output setting

Change lower range in “Basic settings”, see fi g. 35, p. 26

WO15

Life time expired. OXY 1100 should be replaced as soon as
possible

Replace OXY 1100

WO20

The temperature of OXY 1100 is below 0°C (32°F).
The accuracy might be reduced.

Accuracy reduced

WO21

The temperature of OXY 1100 is above 40°C (104°F).
The accuracy might be reduced.

Accuracy reduced

EO50

Calibration not possible.
OXY 1100 is worn or defective

Replace OXY 1100

EO51

Output signal from OXY 1100 is too low.
OXY 1100 is worn or defective

Replace OXY 1100

EO55

The OXY 1100 membrane has been damaged

Replace OXY 1100

EO60

Error in EEPROM

Replace transmitter

EO61

Error in communication between the µ-controller and the A/D
converter or error in the A/D converter

Replace transmitter

EO62

Error in the input circuit

Replace transmitter

EO63

Error in the temperature circuit

Replace transmitter