Symptom

Current signal

Fault

Remedy

Constant current signal
(varying oxygen concentration)

I = 0 mA

Incorrect wiring

Check wiring and make the necessary corrections.

Undervoltage

Check the supply voltage at transmitter.
Supply voltage must be at least 12 V.

Initialisation fault

1. Switch off transmitter for min. 5 seconds.
2. Switch on again while sensor points downwards.

Transmitter defective

Replace transmitter

I = 3.80 mA or
24.0 mA

Initialisation fault

1. Switch off transmitter for min. 5 seconds.
2. Switch on again while sensor points downwards.

Transmitter defective

Replace transmitter

I = 3.85 mA or
23.5 mA

Membrane leakage

Replace OXY 1100 sensor.

Moisture behind OXY 1100 sensor

Fully dry the terminals at transmitter.
A hair dryer might be of assistance.

I = 3.90 mA or
23.0 mA

Calibration not possible, worn OXY 1100 sensor

Replace OXY 1100 sensor.

I = 3.95 mA

Oxygen under set measuring range

Set the lower measuring limit to a lower value.

I = 22.00 mA

Oxygen level higher than set measuring range

Set the upper measuring limit to a higher value.

Error in measuring value

Variable

Calibration value incorrect

1. Check FSO in atmospheric air. Hold the transmitter up in

the air with the sensor downwards. Check value against
DS/EN 25814 (see appendix IIa)

2. Check zero point (see appendix IIb)
3. In the case of large reading error, recalibrate.

Troubleshooting guide - OXY 4100/3150/4150 stand alone transmitter