Forced current and relay outputs, Fig. 49, Service mode – Hach-Lange EVITA OXY User Manual User Manual

Page 40

Advertising
background image

38

In the “service mode” it is possible to control the
relay and the current outputs directly. This may be
useful when e.g. a control loop is to be adjusted or if
an alarm signal should be checked.

Service

mode

USC

I Out 2

USC

I Out 3

USC

I Out 4

USC

Relay 1

USC

Relay 2

USC

Relay 3

USC

Relay 4

USC

Relay 5

USC

Digital input

High

USC

(High/Low)

Mode

Normal

I OUT 1

Normal

Forced

Forced value

00.0 mA

I OUT 1

00.0-24.0 mA

Mode

Normal

RELAY 1

Normal

Open

Closed

Event log

USC

Log xxxxdxxh

xxxxxxxxxxxxxxxx

USC

I Out 1

USC

081R9406.02.02

HART comm. error

0856 0007 001% 00

USC

Fig. 49

Forced current and relay outputs

Service mode

The quality of the HART

®

communication is

displayed in the “HART comm. error” submenu.
The quality is shown as:

VVVV XXXX YYY % ZZ

where

VVVV is the number of received messages

without any errors,

XXXX is the number of received messages

with errors,

YYY % is the percentage of messages that

had errors

ZZ

is the number of messages that were
eliminated, i.e. max. number of retries
was reached.