Forced current and relay outputs, Fig. 49, Service mode – Hach-Lange EVITA OXY User Manual User Manual
Page 40

38
In the “service mode” it is possible to control the
relay and the current outputs directly. This may be
useful when e.g. a control loop is to be adjusted or if
an alarm signal should be checked.
Service
mode
USC
I Out 2
USC
I Out 3
USC
I Out 4
USC
Relay 1
USC
Relay 2
USC
Relay 3
USC
Relay 4
USC
Relay 5
USC
Digital input
High
USC
(High/Low)
Mode
Normal
I OUT 1
Normal
Forced
Forced value
00.0 mA
I OUT 1
00.0-24.0 mA
Mode
Normal
RELAY 1
Normal
Open
Closed
Event log
USC
Log xxxxdxxh
xxxxxxxxxxxxxxxx
USC
I Out 1
USC
081R9406.02.02
HART comm. error
0856 0007 001% 00
USC
Fig. 49
Forced current and relay outputs
Service mode
The quality of the HART
®
communication is
displayed in the “HART comm. error” submenu.
The quality is shown as:
VVVV XXXX YYY % ZZ
where
VVVV is the number of received messages
without any errors,
XXXX is the number of received messages
with errors,
YYY % is the percentage of messages that
had errors
ZZ
is the number of messages that were
eliminated, i.e. max. number of retries
was reached.