52 hart, Fig. 51 – Hach-Lange EVITA OXY User Manual User Manual
Page 54

52
HART
®
The OXY 4100/4150/3150 offers additional
features which can be accessed through the
USC 5000/6000/7000 signal converter or HART
®
communication. To use HART
®
communication it
is necessary to use a PC or PLC with a HART
®
modem and associated software.
All Universal Commands are fully implemented, as
are the necessary Common Practice Commands,
but to ensure complete utilisation of the transmitter,
special Device Specifi c Commands are used.
HART
®
communication makes it possible to read off:
· oxygen concentration
· temperature
· remaining lifetime of the OXY 1100 sensor
· alarms
· indication of completed calibration
When calibrating/replacing an OXY 1100 sensor
the following can be performed with HART
®
communication:
-
initiation of calibration
-
keying in of values for relative humidity, air
pressure and salinity for compensation purposes
-
zero setting of life counter
The following programming is possible with HART
®
communication:
–
setting of measuring range
–
deactivation of tilt-switch for initiating an
automatic calibration
–
setting of current output signal during calibration
for 4 mA, 20 mA or last measuring value
–
deactivation of alarms on current output
–
measurement start after completed calibration
–
change OXY 1100 sensor type
The USC 5000/6000/7000 can act as either a primary
master or a secondary master. This can be set as
shown in fi g. 51.
Fig. 51
USC
HART Settings
USC
Change HART
Trm. Mode
USC
USC
USC
USC
081R9403.02.0
2
Scan HART bus
USC
Scan HART bus
Are you sure ?
USC
Set mode to
Secondary master
Curr. Trm. mode
Primary master
Curr. Trm. mode
Secondary Master
Set mode to
Primary Master
Scan HART bus
busy
USC
Trm. address 0
changed to 05
USC
USC is Primary Master
USC is Secondary Master
USC restarts
USC restarts