52 hart, Fig. 51 – Hach-Lange EVITA OXY User Manual User Manual

Page 54

Advertising
background image

52

HART

®

The OXY 4100/4150/3150 offers additional
features which can be accessed through the
USC 5000/6000/7000 signal converter or HART

®

communication. To use HART

®

communication it

is necessary to use a PC or PLC with a HART

®

modem and associated software.

All Universal Commands are fully implemented, as
are the necessary Common Practice Commands,
but to ensure complete utilisation of the transmitter,
special Device Specifi c Commands are used.

HART

®

communication makes it possible to read off:

· oxygen concentration
· temperature
· remaining lifetime of the OXY 1100 sensor
· alarms
· indication of completed calibration

When calibrating/replacing an OXY 1100 sensor
the following can be performed with HART

®

communication:

-

initiation of calibration

-

keying in of values for relative humidity, air
pressure and salinity for compensation purposes

-

zero setting of life counter

The following programming is possible with HART

®

communication:

setting of measuring range

deactivation of tilt-switch for initiating an
automatic calibration

setting of current output signal during calibration
for 4 mA, 20 mA or last measuring value

deactivation of alarms on current output

measurement start after completed calibration

change OXY 1100 sensor type

The USC 5000/6000/7000 can act as either a primary
master or a secondary master. This can be set as
shown in fi g. 51.

Fig. 51

USC

HART Settings

USC

Change HART

Trm. Mode

USC

USC

USC

USC

081R9403.02.0

2

Scan HART bus

USC

Scan HART bus

Are you sure ?

USC

Set mode to
Secondary master

Curr. Trm. mode
Primary master

Curr. Trm. mode
Secondary Master

Set mode to
Primary Master

Scan HART bus

busy

USC

Trm. address 0

changed to 05

USC

USC is Primary Master

USC is Secondary Master

USC restarts

USC restarts