25

USC 7000 multidrop system

Changing transmitter address from default

To allow the USC 7000 to recognise the transmitter
on the HART

®

bus, the transmitter needs a unique

address between 1 and 15. The EVITA

®

OXY

transmitter is factory delivered with the address 0,
which is used for point-to-point installations (USC
5000/6000).

To programme the address of a new transmitter,
connect the transmitter to the USC 7000 and press
“Scan HART bus”. The transmitter will get the next
available address between 1 and 15.

Note: If, before scanning the HART

®

bus, a trans-

mitter with address higher than 0 was registered as
interrupted, the new transmitter will be assigned this
address.

Rem. lifetime

Transmitter 0

72. h

No. of alarms

Transmitter 0

12. d

Calibration

Transmitter 0

42. d

Temperature

Transmitter 0

Oxygen

Transmitter 0

10.64 mg/l

Error pending

Scan HART bus

Password

0000.

Settings

USC

Basic settings

TR. 0

Advanced settings

TR. 0

Settings OXY

Ser.No. SSSSVVNUUÅ

TR. 0

Set transmitter
address 00

TR. 0

081R9302.02.02

Reset lifetime

Transmitter 0

Last calibration

Transmitter 0

USC

Scan HART bus

Are you sure ?

USC

Scan HART bus

Busy

USC

Trm. address 0

changed to 05

Fig. 34