Usc 7000 multidrop system, Changing transmitter address from default, Fig. 34 – Hach-Lange EVITA OXY User Manual User Manual
Page 27

25
USC 7000 multidrop system
Changing transmitter address from default
To allow the USC 7000 to recognise the transmitter
on the HART
®
bus, the transmitter needs a unique
address between 1 and 15. The EVITA
®
OXY
transmitter is factory delivered with the address 0,
which is used for point-to-point installations (USC
5000/6000).
To programme the address of a new transmitter,
connect the transmitter to the USC 7000 and press
“Scan HART bus”. The transmitter will get the next
available address between 1 and 15.
Note: If, before scanning the HART
®
bus, a trans-
mitter with address higher than 0 was registered as
interrupted, the new transmitter will be assigned this
address.
Rem. lifetime
Transmitter 0
72. h
No. of alarms
Transmitter 0
12. d
Calibration
Transmitter 0
42. d
Temperature
Transmitter 0
Oxygen
Transmitter 0
10.64 mg/l
Error pending
Scan HART bus
Password
0000.
Settings
USC
Basic settings
TR. 0
Advanced settings
TR. 0
Settings OXY
Ser.No. SSSSVVNUUÅ
TR. 0
Set transmitter
address 00
TR. 0
081R9302.02.02
Reset lifetime
Transmitter 0
Last calibration
Transmitter 0
USC
Scan HART bus
Are you sure ?
USC
Scan HART bus
Busy
USC
Trm. address 0
changed to 05
Fig. 34