Hach-Lange EVITA OXY User Manual User Manual

Page 51

Advertising
background image

49

Disposal in accordance with EU directive 2002/96/EC

In compliance with EU directive 2002/96/EC, HACH LANGE / the local subsidiary will accept back and dispose
of the old instrument at no cost.

ATTENTION!

It is not allowed to dispose of the instrument using municipal waste
disposal services. Please talk to your local HACH LANGE contact.

The worn OXY 1100 sensor must be disposed as ordinary industrial waste, and the worn OXY 4100/4150 transmitter must be
disposed as electronic waste. Before disposing the OXY 4100/4150, the cable must be cut off and disposed as metallic waste.

Important: Before returning goods to HACH LANGE, the products must be cleaned properly.

System accuracy

Digital output:

±0.1 % of FSO at calibrating temperature

Analogue output:

±0.2 % of FSO at calibrating temperature

Digital/analogue output: ±0.5 % of FSO at 0 - 40°C (32°F to 104°F)

Reaction time OXY 1100

25 µm:

τ = 7 s

50 µm:

τ = 22 s

125 µm:

τ = 110 s

Diaphragm leakage monitoring

Measurement between anode and a terminal in contact with the measured medium

Enclosure

OXY 4100:

IEC 529: IP 68 (1 m); NEMA 6P (3’)

OXY 3150/4150:

IEC 529: IP 68 (10 m); NEMA 6P (30’)

Dimensions and weight

OXY 4100:

d = 240 mm (9.6”);

2.7 kg (6 lb)

OXY 3150/4150:

d = 50 mm; l = 180 mm (d = 2.0”; l = 7.2”);

1 kg (2.2 lb)

Materials

PBT/PC

EMC

Emission:

IEC/EN 61000-6-3

Immunity:

IEC/EN 61000-6-2